Area Peak Factors
Empirical
APFs
This menu allows
the user to selectively load empirical APFs (Area Peak Factors)
from a supplied ASCII file or data entry.
A powerful feature
in Probe for Windows allows the user to select an Area-Peak-Factor
(APF) correction for use in correcting x-ray intensities for peak
shift and shape changes between the standard and the unknown. This
can occur especially with Ka lines of the light elements such as
oxygen, nitrogen, carbon and boron. With these elements, differences
in the bond energies between say TiC and SiC will produce significantly
different peak shapes resulting in systematic errors in the analysis
of these materials when only the peak x-ray intensities are used.
For example,
when analyzing oxygen and using MgO or Fe2O3 as an oxygen standard,
analyses of SiO2 will produce errors up to about 4.5%. This effect
is independent of the matrix correction and can be corrected for
only by the use of the appropriate APFs.
The APF concept
was proposed by Bastin and Heijligers (Bastin and Heijligers, 1986)
to provide a simple correction scheme for this effect. It basically
requires measuring the ratio of the peak intensity to the integrated
intensity between a primary standard to a number of secondary standards
on the light element spectrometer used. The wavelength scan feature
in Probe for Windows can be used to acquire the peak shape profiles.
After correcting for background and removal of extraneous peaks
from interfering lines, the APF can be calculated. The calculation
of the APF factor is shown below :

Where :
is the integrated intensity of the secondary standard
is the peak intensity of the primary standard
is the peak intensity of the secondary standard
is the integrated intensity of the primary standard
The peak shape
and shift of light element peaks can be easily demonstrated using
MgO and SiO2 standard samples. First select a suitable layered dispersive
element (LDE) analyzing crystal for oxygen Ka analysis. Tune the
spectrometer using the MgO standard and acquire a standard sample
for MgO. Assign MgO as the calibration standard for oxygen. Then
move to the SiO2 standard and acquire another quantitative standard
sample on the SiO2 sample. Note that the analysis of SiO2 using
MgO as a calibration standard will result in a low total for the
SiO2, somewhere around 96-97 %. This is not an error in the mass
absorption coefficient (MAC), and no attempt should be made to correct
this problem by entering an arbitrary MAC value.
Instead, the
proper APF correction will need to be applied as shown below. Although
the values below are relative to Fe2O3, you will note that the APF
for MgO is 1.000, which means that these correction factors apply
equally well relative to MgO. In the Oxygen Area Peak Factor table
below you will note that oxygen in the presence of SiO2 needs to
be corrected by a factor of 1.0444 when MgO or Fe2O3 is used a a
calibration standard. After entering this APF correction factor
for the oxygen channel and re-analyzing, note that the total is
now close to 100 %.
It is very important
to note that the APF values selected or entered are always measured
relative to some standard sample. For example, if measuring C Ka
and using Fe3C as the primary standard for carbon, then any C Ka
APF values used must be those measured relative to Fe3C. For the
same reason, if using APF values for a particular (light) element
and one decides to re-assign the standard for that element, the
APF values for that element must also be changed to reflect the
standard re-assignment. See the section above for information on
re-calculating the APF values relative to another standard.
For the above
reason it is usually most efficient to simply always use the same
standard for each light element analyzed. Typically (in order to
utilize the APF values in the supplied EMPAPF.DAT file) these will
be :
- Oxygen : MgO
or Fe2O3
- Nitrogen : AlN
- Carbon : Fe3C
- Boron : B metal
The APF correction
in Probe for Windows will allow the user to enter empirical APF
values in each run. The user may enter one or more APF factors for
each emitting element although they are generally applied to soft
x-ray lines. Note however that even S Ka exhibits peak shift and
shape changes when comparing sulfide and sulfate peaks. The APF
for each absorber will be summed according to it's weight fraction
in the composition and applied to the emitting element counts during
the ZAF or Phi-Rho-Z iteration.
If it is desired
to correct a complete unknown for peak shape changes, simply perform
a wavelength scan on the standard being used for the light element
emitter (oxygen, carbon, etc.) and also on the unknown. Calculate
the APF using the expression above and from the Analytical | Empirical
APFs menu, enter the same APF for all absorbers in the unknown.
The program will then sum the APFs which will be the same as the
single measured APF. Note that this APF may only apply to a single
composition. One can perform additional wavelength scans on other
samples to confirm this however.
The APF correction
values are defined in the EMPAPF.DAT file in the XRAYDATA sub directory
(usually C:\PROBEWIN\XRAYDATA). The file contains some 50 or 60
values that may or may not be applicable, depending on the analyzing
crystals and standards available. The user may edit the file using
any ASCII text editor such as NotePad or NoteBook to insert their
own measurements. Be careful to avoid adding any <tab> characters
when editing the file (use space or comma delimited values only).
Note that the elements may be in any order although they have been
sorted to facilitate editing.
The format of
the EMPAPF.DAT file is shown below :
"b"
"ka" "c" 1.02 "B4C/B/STE"
"b" "ka" "n" 1.2 "BN/B/STE"
"b" "ka" "al" 1.12 "AlB2/B/STE"
"b" "ka" "al" 1.01 "AlB12/B/STE"
"b" "ka" "si" 1 "SiB3/B/STE"
"b" "ka" "si" .92 "SiB6/B/STE"
"b" "ka" "ti" .75 "TiB/B/STE"
"b" "ka" "ti" .88 "TiB2/B/STE"
"b" "ka" "v" 1 "VB2/B/STE"
"b" "ka" "cr" .9 "CrB/B/STE"
"b" "ka" "cr" 1.1 "CrB2/B/STE"
"b" "ka" "fe" 1.1 "FeB/B/STE"
"b" "ka" "fe" 1.25 "Fe2B/B/STE"
"b" "ka" "co" 1.2 "CoB/B/STE"
"b" "ka" "co" 1.02 "Co2B/B/STE"
"b" "ka" "ni" 1.2 "NiB/B/STE"
"b" "ka" "ni" 1.06 "Ni2B/B/STE"
"b" "ka" "ni" .98 "Ni3B/B/STE"
"b" "ka" "zr" .8 "ZrB2/B/STE"
"b" "ka" "nb" .8 "NbB/B/STE"
"b" "ka" "nb" .9 "NbB2/B/STE"
"b" "ka" "mo" .94 "MoB/B/STE"
"b" "ka" "la" .9 "LaB6/B/STE"
"b" "ka" "ta" .88 "TaB/B/STE"
"b" "ka" "ta" 1.1 "TaB2/B/STE"
"b" "ka" "w" .98 "WB/B/STE"
"b" "ka" "u" 1.04 "UB4/B/STE"
"c" "ka" "b" 1.01 "B4C/Fe3C/WSi/59.8"
"c" "ka" "si" .933 "SiC/Fe3C/WSi/59.8"
"c" "ka" "ti" .868 "TiC/Fe3C/WSi/59.8"
"c" "ka" "v" .873 "V2C/Fe3C/WSi/59.8"
"c" "ka" "v" .873 "VC/Fe3C/WSi/59.8"
"c" "ka" "cr" .8 "Cr7C3/Fe3C/STE"
"c" "ka" "cr" .83 "Cr3C2/Fe3C/STE"
"c" "ka" "cr" .8 "Cr23C6/Fe3C/STE"
"c" "ka" "zr" .88 "ZrC/Fe3C/WSi/59.8"
"c" "ka" "nb" .79 "NbC/Fe3C/STE"
"c" "ka" "mo" .82 "Mo2C/Fe3C/STE"
"c" "ka" "hf" .83 "HfC/Fe3C/STE"
"c" "ka" "ta" .96 "TaC/Fe3C/STE"
"c" "ka" "w" .97 "WC/Fe3C/STE"
"c" "ka" "w" 1.02 "W2C/Fe3C/STE"
"n" "ka" "si" 1.103 "Si3N4/AlN/WSi/59.8"
"n" "ka" "ti" .997 "TiN/AlN/WSi/59.8"
"n" "ka" "v" 1.0226 "VN/AlN/WSi/59.8"
"n" "ka" "cr" 1.018 "Cr2N/AlN/WSi/59.8"
"n" "ka" "fe" 1.012 "Fe2N/AlN/WSi/59.8"
"n" "ka" "zr" .9952 "ZrN/AlN/WSi/59.8"
"n" "ka" "hf" 1.002 "HfN/AlN/WSi/59.8"
"o" "ka" "b" 1.0628 "B6O/Fe2O3/WSi/59.8"
"o" "ka" "mg" 1 "MgO/Fe2O3/WSi/59.8"
"o" "ka" "al" 1.0213 "Al2O3/Fe2O3/WSi/59.8"
"o" "ka" "si" 1.0444 "SiO2/Fe2O3/WSi/59.8"
"o" "ka" "ti" .9796 "TiO2/Fe2O3/WSi/59.8"
"o" "ka" "cr" .993 "Cr2O3/Fe2O3/WSi/59.8"
"o" "ka" "mn" 1.0121 "MnO/Fe2O3/WSi/59.8"
"o" "ka" "fe" .9962 "Fe3O4/Fe2O3/WSi/59.8"
"o" "ka" "co" 1.0133 "CoO/Fe2O3/WSi/59.8"
"o" "ka" "ni" 1.0153 "NiO/Fe2O3/WSi/59.8"
"o" "ka" "cu" .9946 "Cu2O/Fe2O3/WSi/59.8"
"o" "ka" "cu" .9943 "CuO/Fe2O3/WSi/59.8"
"o" "ka" "zn" .9837 "ZnO/Fe2O3/WSi/59.8"
"o" "ka" "ga" 1 "Ga2O3/Fe2O3/WSi/59.8"
"o" "ka" "zr" .9823 "Y3Fe5O12/Fe2O3/WSi/59.8"
The first column
(in the example is "b" for Boron) is the atomic symbol
of the emitting element. The second column ("ka" in the
example equals Ka) is the x-ray line of the emitter. The third column
is the atomic symbol of the absorber element. The fourth column
is the actual area peak factor (APF) that has been experimentally
measured. The last parameter is a string that contains the conditions
under which the APF was measured. For example, the comment "B4C/B/STE"
indicates that the peak and integrated intensities for the APF were
measured on a B4C (boron carbide) sample, relative to a B (elemental
Boron) primary standard, using a Pb Stearate analyzer. The reason
that this information is needed is because the APF is a relative
measurement (to a standard) and the value is dependent on the resolution
(crystal type) used. If a LDE (layered dispersive element) analyzer
is used, the 2d of the analyzer should be indicated also as seen
in the last two lines of the example.
The first and
third columns (emitting atomic symbol and absorber atomic symbol)
must be valid element symbols. The second column (x-ray line) must
be "ka", "la" or "ma". All element
and x-ray symbols must be enclosed in double quotes. The fourth
column (APF value) must be a real number greater than 0.0. The comment
string must be enclosed in double quotes.
These values
can then be loaded from the Analytical | Empirical APFs menu item
in Probe for Windows. Note that the comments listed after each APF
value can be used to indicate the experimental conditions under
which the APF was measured. For consistency sake, the formula of
the compound measured is listed first, then the primary reference
standard used and finally the analyzing crystal type (and 2d spacing
if desired) used. Edit the EMPAPF.DAT file in the XRAYDATA sub directory
to add additional APFs from your own measurements.
The following
is a short listing of some typical APFs taken from Bastin and Heijligers
(1986). Because the factors are spectrometer and crystal dependent
these should be used as a guide only.
Oxygen
Area Peak Factors (APF)
Relative
to Fe2O3 or MgO using W/Si LDE (2d = 59.8)
Oxide
|
Area-Peak
Factor (APF)
|
B6O
|
1.0628
|
MgO
|
1.0000
|
Al2O3
|
1.0213
|
SiO2
|
1.0444
|
TiO2
|
0.9796
|
Cr2O3
|
0.9930
|
Mn3O4
|
1.0121
|
Fe3O4
|
0.9962
|
CoO
|
1.0133
|
NiO
|
1.0153
|
Cu2O
|
0.9946
|
CuO
|
0.9943
|
ZnO
|
0.9837
|
Ga2O3
|
1.0000
|
Y3Fe5O12
|
0.9823
|
Carbon
Area-Peak Factors
Relative
to Fe3C using W/Si LDE (2d = 59.8):
Carbide
|
Area-Peak-Factor
(APF)
|
B4C
|
1.010
|
SiC
|
0.933
|
TiC
|
0.868
|
V2C
|
0.873
|
VC
|
0.873
|
ZrC
|
0.880
|
Nitrogen
Area-Peak Factors
Relative
to AlN using W/Si LDE (2d = 59.8):
Carbide
|
Area-Peak-Factor
(APF)
|
AlN
|
1.000
|
Si3N4
|
1.103
|
TiN
|
.997
|
VN
|
1.0226
|
Cr2N
|
1.018
|
Fe2N
|
1.012
|
ZrN
|
.9952
|
HfN
|
1.002
|
Area
Peak Factors and Standard Assignments
Another consideration
in assigning standards, concerns the use of Area Peak Factors (APF)
for correction of light element peak shift and shape changes. When
analysis of light elements is to be performed (typically oxygen,
nitrogen, carbon and boron but possibly heavier elements such as
silicon, aluminum and sulfur), you need to plan the use of APF corrections
for the most accurate results. The actual APF values selected will
depend the standard assigned for calibration of the light element.
For example,
when analyzing for C Ka and using Fe3C as your carbon standard,
your APF value for C Ka in a Cr7C3 matrix might be 0.80. However,
if instead, you re-assigned your carbon standard to TiC, you must
use APF values relative to TiC rather than Fe3C. If they are available
in the Empirical APF database (EMPAPF.DAT), just change the APF
values for C Ka. Otherwise you would need to calculate the Fe3C
APF values relative to TiC. Some typical Fe3C APF values for C Ka,
recalculated relative to TiC and Cr7C3, are shown in the table below
:
APF
Relative To : |
Fe3C
(APF/1.000) |
TiC
(APF/0.868) |
Cr7C3
(APF/0.80) |
Fe3C
|
1.000
|
1.152
|
1.25
|
TiC
|
0.868
|
1.000
|
1.085
|
Cr7C3
|
0.80
|
0.92
|
1.00
|
G. F. Bastin
and H. J. M. Heijligers, "Quantitative Electron Probe Microanalysis
of Carbon in Binary Carbides," Parts I and II, X-Ray Spectr.
15: 135-150, 1986
|